Ablikim, U.Zohrabi, M.Jochim, B.Berry, B.Severt, T.Carnes, K. D.Ben-Itzhak, Itzik2016-04-062016-04-06http://righttp://hdl.handle.net/2097/32449Citation: Ablikim, U., Zohrabi, M., Jochim, B., Berry, B., Severt, T., Carnes, K. D., & Ben-Itzhak, I. (2015). Note: Position dependence of time signals picked off a microchannel plate detector. Review of Scientific Instruments, 86(1), 3. doi:10.1063/1.4906327Using an ultrafast laser and a precision mask, we demonstrate that time signals picked off directly from a microchannel plate detector depend on the position of the hit. This causes a time spread of about 280 ps, which can affect the quality of imaging measurements using large detectors. (C) 2015 AIP Publishing LLC.Copyright 2015 AIP Publishing. This article may be downloaded for personal use only. Any other use requires prior permission of the author and AIP Publishing. The following article appeared in U. Ablikim, M. Zohrabi, Bethany Jochim, B. Berry, T. Severt, K. D. Carnes and I. Ben-Itzhak., Rev. Sci. Instrum. 86, 016111 (2015) and may be found at http://dx.doi.org/10.1063/1.4906327.Momentum SpectroscopyGenerationEfficiencyAnodeInstruments & InstrumentationPhysicsNote: Position dependence of time signals picked off a microchannel plate detectorArticle