Note: Position dependence of time signals picked off a microchannel plate detector

dc.citation.doi10.1063/1.4906327
dc.citation.issn0034-6748
dc.citation.issue1
dc.citation.jtitleReview of Scientific Instruments
dc.citation.volume86
dc.contributor.authorAblikim, U.
dc.contributor.authorZohrabi, M.
dc.contributor.authorJochim, Bethany
dc.contributor.authorBerry, B.
dc.contributor.authorSevert, T.
dc.contributor.authorCarnes, K. D.
dc.contributor.authorBen-Itzhak, I.
dc.date.accessioned2023-12-07T22:33:53Z
dc.date.available2023-12-07T22:33:53Z
dc.date.issued2015-01-20
dc.date.published2015-01-20
dc.description.abstractUsing an ultrafast laser and a precision mask, we demonstrate that time signals picked off directly from a microchannel plate detector depend on the position of the hit. This causes a time spread of about 280 ps, which can affect the quality of imaging measurements using large detectors.
dc.identifier.urihttps://hdl.handle.net/2097/43940
dc.relation.urihttps://doi.org/10.1063/1.4906327
dc.rightsThis article may be downloaded for personal use only. Any other use requires prior permission of the author and AIP Publishing. This article appeared in [Note: Position dependence of time signals picked off a microchannel plate detector. Review of Scientif
dc.rights.urihttp://rightsstatements.org/vocab/InC/1.0/
dc.rights.urihttp://web.archive.org/web/20180624131647/https://publishing.aip.org/authors/web-posting-guidelines
dc.titleNote: Position dependence of time signals picked off a microchannel plate detector